AirSPEC
Smart gamma spectrometer - radiometer

Spectrometer AirSPEC is intended for automated radiation environment monitoring, measurements of ambient air volumetric activity  and gamma-radiation ambient dose equivalent rate  nearby the filter, comprising natural and artificial gamma-radiating radionuclides detected in aerosoles of ambient air surface layer.

The spectrometer is applied in the systems of air sampling as well as radiation monitoring stations.

Application

Smart gamma spectrometer AirSPEC

Description

The spectrometer AirSPEC is a all-in-one unit which contains processing unit within the cover. Smart detection unit of the spectrometer contains scintillation detector (several type), photomultiplier tube (PMT), spectrometric tract stabilization system (LED or Am-241), HV-generator, amplifier, multichannel pulses analyzer (MCA) and temperature sensors. The spectrometer is placed into thermostabilized and damp proof housing.

Though the present measurement system is autonomous and automated, the operator thanks to the possible remote access to the spectrometer may perform the analysis of the measured spectra with application of modern and comprehensive software.

Composition

 - Spectrometer AirSPEC
 - Thermostabilized housing (TSH-L or TSH-Pro)
 - Connective cables set
 - Power supply adapter
 - Spectrometric analysis software
 - User manual for Spectrometer AirSPEC

Specification

Detector type
AirSPEC-N76
AirSPEC-L51
AirSPEC-C51
AirSPEC-S51


NaI(Tl) Ø76x76 mm
LaBr3(Ce) Ø51x51 mm
CeBr3 Ø51x51 mm
SrI2(Eu) Ø38x38 mm

The energy range of measuring distribution of gamma radiation :

30 to 3000 keV

Relative energy resolution on the line 661.7 keV, measuring with
radionuclide source Cs-137, %, not more than
AirSPEC-N76
AirSPEC-L51
AirSPEC-C51
AirSPEC-S51



8
3.3
4
3.5

Integral nonlinearity in the gamma energy range from 40 to 3000 keV,%, not more then

±1%

Range of the measurements of the dose rate , µSv/h

0.01 - 3000

Maximum throughput, kcps, not less than

100

Channels quantity in the spectrum

1024,2048,4096,8192,16384

Data transfer interfaces

Ethernet, RS-485, USB

Operation conditions in thermostabilized housing
• Ambient air temperature
for termostabilized housing TSH-L
for termostabilized housing TSH-Pro
• Air relative humidity
• Atm. pressure within the range, kPа
• Intensity of magnetic fields of the permanent and variable grid frequencies



-40 ... +40 °С
-40 ... +60 °С
up to 100 %
84 ÷ 106.7
40 А/m

Overall dimensions (standard variant AirSPEC-N76), mm :
- Spectrometer
- Housing


Ø93x340
Ø160x440

Shell protection level

IP68

Original calibration data set

Ra-226, Th-232, K-40, Cs-137, Co-60