PM1610
Gamma and X-Ray personal dosimeter

Application

Electronic personal dosimeters of PM1610 series provide measurement of ambient dose equivalent and ambient dose equivalent rate of continuous and pulsed gamma and X-ray radiation in the energy range from 20 keV to 10 MeV, as well as time of dose accumulation. Instruments have visual, auido and vibration alarms to warn the user when preset dose(dose rate) threshold levels are exceeded. Shockproof rubberized case and  LCD with fluorescent backlight allows operation in harsh environment. Dosimeters are compact and easy for use, have long life built-in rechargeable battery.

PM1610

Network functions

Personal Dose Tracker System developed by Polimaster, enables connection up to 100 units of PM1610.  PDT System is intended for continuous account for external radiation exposure, maintenance of dose history database and calculation of the personnel accumulated doses (individual as well as group doses) during the set period of time, alarm when preset thresholds (individual as well as group thresholds) are exceeded. Software enables information transfer by USB-protocol to a personal computer to analyze, process, and generate corresponding databases in a control center or an expert center.

Description

Dosimeters allow setting of two dose rate alarm threshold levels and two dose alarm threshold levels. Instruments continuously control the threshold levels and alert the user with audible, light and vibration alarms when the threshold levels are exceeded. Instruments automatically record data into permanent memory and store up to 7500 events of dose rate and dose measurement history in non-volatile memory. PM1610-01 modification is equipped with data communication channel (ISO15693 standard).

Specification

Detector type

Geiger-Muller tube

Ambient equivalent dose rate of gamma and X radiation measurement range

0.1 µSv/h...10 Sv/h

Ambient equivalent dose of gamma and X radiation measurement range:
∙ PM1610, PM1610-01:
​ - continuous photon radiation
- pulsed photon radiation
∙ PM1610A, PM1610A-01:
​ - continuous photon radiation
- pulsed photon radiation




0.05 µSv ... 10.0 Sv
10 µSv ... 10.0 Sv

​0.05 µSv ... 20.0 Sv
10 µSv ... 20.0 Sv

Relative error of dose rate measurement in the range from 0.1 µSv/h to 10 Sv/h:
- PM1610, PM1610-01
- PM1610A, PM1610A-01



​±(15+0.0015/Ḣ)% (where Ḣ - dose rate, mSv/h)
± (10+K1/Ḣ+К2∙Ḣ )% (where Ḣ - dose rate, mSv/h,
K1 - coefficient 0.0015 mSv/h,
К2 - coefficient 0.0015 (mSv/h)^-1)

Gamma and X-ray radiation energy range

20 keV...10 MeV

Energy response relative to 0.662 MeV:
- from 20 to 33 keV
- from 33 to 48 keV
- from 48 keV to 3 MeV
- from 3 to 10 MeV


60 %
40 %
30 %
50 %

PC communication

USB

Power supply

built-in rechargeable battery
(502030 240mAh MOBILE STAR)

Battery lifetime: average DER not more than 0.3 µSv/h, backlight, audible, light and vibration alarms activated for not more than 20 s/24 hours1 month

1 month

Drop test on concrete floor

1.5 m

Dimensions

58 x 59 x 20 mm

Weight

70 g

Alarm Type

visual, audio, vibration

Operating temperature
Humidity
Atmospheric pressure

-20 °С to + 50 °С
98 % at 35 °С
84 to 106.7 kPa

Protection class

​IP65

Standards compliance

Designed to meet ANSI N42.20 - 2006 and IEC 61526: 2010 requirements

Complies with ANSI 42.34, section 9.11 requirements by the level of emitted radio-noise